Banner Length and Dimensional Calibration Laboratory

Banner Length and Dimensional Calibration Laboratory

1. Banner 1 LDC-RCLC
2. Banner 1 LDC-RCLC
3. Banner 1 LDC-RCLC
previous arrow
next arrow

Fig 1. Laser Head.

Fig 2. Laser distance meter.

Fig 3. XAMPP Software.

Fig 4. The design of smart measuring probe (SMP)

Fig 5. Laser distance meter measuring bench 20 m

The significant product developed is a Smart Measuring Probe (SMP) designed for calibration of Laser Distance Meters (LDM). It features:
โ€ข Automatic high-precision positioning (repeatability ยฑ0.002 mm).
โ€ข Wireless DC motor control.
โ€ข Integration with interferometry system for reduced measurement uncertainty.
โ€ข Complies with ISO 16331-1:2012.
The SMP enables accurate, automated calibration of LDMs up to 20 meters, reducing human errors and ensuring traceability with measurement uncertainties as low as 0.64 mm.


  • Developing and improving calibration systems for length measurement.
  • Conducting research in advanced laser metrology.
  • Providing calibration services for LDMs and other distance instruments.
  • Supporting national metrology traceability and international standards compliance.

  • Laser distance metrology and calibration.
  • Development of interferometry systems.
  • Evaluation and reduction of measurement uncertainties.
  • Smart automation for calibration instruments.
  • Smart Measuring Probe: SMP

This intelligent probe is specially designed for LDM calibration system. The body is made of high grade aluminum (7075), the movement is controlled by high precision DC motor, wireless control, the control software is XAMPP, which can set the measuring points automatically and collect the data from CCD camera until all the calibration points are reached.


  • Laser Distance Meters Calibration base on ISO 16331-1 (2012).
  • Calibration points (CP) up to 10CP (following ISO 16331-1 (2012).
  • Laboratory conditions with
    Temperature 25 ยฑ3ยฐC
    Humidity 50 ยฑ10%
    Target refectivity 95ยฑ5%
    Background illumination < 3000 lx
  • Measurement uncertainties
    This measurement uncertainty evaluation based on M3003 (2012)
    Standard accuracy
    Dead path error of standard
    Resolution of standard
    Resolution of UUC
    Abbe error of measuring bench
    Abbe error of standard
    Cosine error of standard
    Cosine error of UUC
    Drift of standard
    Refractive index of air
    Zero positioning error of standard
    Temperature effect
    Deviation

Laser Distance Meter

  • 0 m to 6 m *CMC 0.25 mm
  • > 6 m to 10 m *CMC 0.36 mm
  • > 10 m to 14 *CMC 0.50 mm
  • > 14 m to 20 m *CMC 0.70 mm

Base on ISO 16331-1 : 2012
CMC = Calibration and Measurement Capability


  • Smart Measuring Probe (SMP)
  • Laser measurement system 5529A (He-Ne Laser 633 nm)
  • CCD camera with data capture software (XAMPP-based)
  • Calibration Bench up to 20 m

  • Accredited with ISO/IEC 17025
  • Developed one of the first automated LDM calibration systems in Thailan


Mr. Pisit Hormchoey
Position : Head of Length and Dimensional Calibration Group (Scientist, Professional Level)
Email : pisit@dss.go.th
Telephone : 0 2201 7058

Mr. Tanawut Sukreetha
Position : Scientist, Practitioner Level
Email : stanawut@dss.go.th
Telephone : 0 2201 7058

Mr. Ananya Mingmueang
Position : Scientist, Practitioner Level
Email : ananya@dss.go.th
Telephone : 0 2201 7058

Review Your Cart
0
Add Coupon Code
Subtotal

 
Scroll to Top